(0018,1083) Intervals Acquired
2C
Number of R-R intervals acquired. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and Cardiac Synchronization Technique (0018,9037) equals other than NONE. Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and Cardiac Synchronization Technique (0018,9037) equals other than NONE. CARDIAC SYNCHRONIZATION MODULE ATTRIBUTES
(0018,1083) >>Intervals Acquired
3
Number of heartbeats that fall within Low R-R Value (0018,1081) and High R-R Value (0018,1082), and were therefore accepted and contribute gamma events to this R-R Interval. NM MULTI-GATED ACQUISITION MODULE ATTRIBUTES
(0018,1083) Intervals Acquired
3
Number of R-R intervals acquired. MR IMAGE MODULE ATTRIBUTES
(0018,1083) Intervals Acquired
3
Number of heartbeats that fall within Low R-R Value (0018,1081) and High R-R Value (0018,1082), and were therefore accepted and contribute coincidence events to this R-R Interval. PET IMAGE MODULE ATTRIBUTES