(0018,1081) |
Low R-R Value |
2C
|
R-R interval low limit for beat rejection, in ms. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and Cardiac Synchronization Technique (0018,9037) equals PROSPECTIVE or RETROSPECTIVE. Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and Cardiac Synchronization Technique (0018,9037) equals PROSPECTIVE or RETROSPECTIVE. |
CARDIAC SYNCHRONIZATION MODULE ATTRIBUTES |
(0018,1081) |
>>Low R-R Value |
3
|
R-R interval lower limit for beat rejection, in msec |
NM MULTI-GATED ACQUISITION MODULE ATTRIBUTES |
(0018,1081) |
Low R-R Value |
3
|
R-R interval low limit for beat rejection, in msec |
US IMAGE MODULE ATTRIBUTES |
(0018,1081) |
Low R-R Value |
3
|
R-R interval low limit for beat rejection, in msec |
MR IMAGE MODULE ATTRIBUTES |
(0018,1081) |
Low R-R Value |
1C
|
R-R interval lower limit for beat rejection, in msec. Required if Series Type (0054,1000), Value 1 is GATED and Beat Rejection Flag (0018,1080) is Y. |