DICOM Tables


MR SPECTROSCOPY FOV/GEOMETRY MACRO ATTRIBUTES

Tag
Tag Name
Requirement
Description
(0018,9103) MR Spectroscopy FOV/Geometry Sequence
1
Identifies the geometry parameters of this frame. Only a single Item shall be permitted in this sequence.
(0018,9127) >Spectroscopy Acquisition Data Columns
1C
Number of data points in the columns direction Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
(0018,9095) >Spectroscopy Acquisition Phase Rows
1C
Number of Phase Encoding Rows Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
(0018,9234) >Spectroscopy Acquisition Phase Columns
1C
Number of Phase Encoding Columns Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
(0018,9159) >Spectroscopy Acquisition Out-of-plane Phase Steps
1C
Number of out-of-plane Phase Encoding steps Required if MR Spectroscopy Acquisition Type (0018,9200) equals PLANE and Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
(0018,0093) >Percent Sampling
1C
Fraction of acquisition matrix lines acquired, expressed as a percent. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.
(0018,0094) >Percent Phase Field of View
1C
Ratio of field of view dimension in phase direction to field of view dimension in frequency direction, expressed as a percent. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.