DICOM Tables


DX POSITIONING MODULE ATTRIBUTES

Tag
Tag Name
Requirement
Description
(0018,5104) Projection Eponymous Name Code Sequence
3
A Sequence that describes the radiographic method of patient, tube and detector positioning to achieve a well described projection or view. Only a single Item shall be permitted in this Sequence. Shall be consistent with the other Attributes in this Module, if present, but may more specifically describe the image acquisition.
(0018,5100) Patient Position
3
Description of imaging subject’s position relative to the equipment. See C.7.3.1.1.2 for Defined Terms and further explanation. If present, shall be consistent with Patient Gantry Relationship Code Sequence (0054,0414) and Patient Orientation Modifier Code Sequence (0054,0412).
(0018,5101) View Position
3
Radiographic view of the image relative to the imaging subject’s orientation. Shall be consistent with View Code Sequence (0054,0220). See C.8.11.5.1.1 for further explanation.
(0054,0220) View Code Sequence
3
Sequence that describes the projection of the anatomic region of interest on the image receptor. Note: It is strongly recommended that this Attribute be present, in order to ensure that images may be positioned correctly relative to one another for display. Shall be consistent with View Position (0018,5101). See C.8.11.5.1.1 for further explanation. Only a single Item shall be permitted in this Sequence.
(0054,0222) >View Modifier Code Sequence
3
View modifier. Zero or more Items may be included in this Sequence.
(0054,0410) Patient Orientation Code Sequence
3
Sequence that describes the orientation of the patient with respect to gravity. See C.8.11.5.1.3 for further explanation. Only a single Item shall be permitted in this Sequence.
(0054,0412) > Patient Orientation Modifier Code Sequence
3
Patient Orientation Modifier. Required if needed to fully specify the orientation of the patient with respect to gravity. Only a single Item shall be permitted in this Sequence.
(0054,0414) Patient Gantry Relationship Code Sequence
3
Sequence which describes the orientation of the patient with respect to the gantry. Only a single Item shall be permitted in this Sequence.
(0018,1111) Distance Source to Patient
3
Distance in mm from source to the table, support or bucky side that is closest to the Imaging Subject, as measured along the central ray of the X-Ray beam. Note: 1. This definition is less useful in terms of estimating geometric magnification than a measurement to a defined point within the Imaging Subject, but accounts for what is realistically measurable in an automated fashion in a clinical setting. 2. This measurement does not take into account any air gap between the Imaging Subject and the “front” of the table or bucky. 3. If the detector is not mounted in a table or bucky, then the actual position relative to the patient is implementation or operator defined. 4. This value is traditionally referred to as Source Object Distance (SOD).
(0018,1110) Distance Source to Detector
3
Distance in mm from source to detector center. Note: This value is traditionally referred to as Source Image Distance (SID). Estimated Radiographic Magnification Factor (0018,1114) 3 Ratio of Source Image Distance (SID) over Source Object Distance (SOD).
(0018,1508) Positioner Type
2
Defined Terms: CARM COLUMN MAMMOGRAPHIC PANORAMIC CEPHALOSTAT RIGID NONE Notes: 1. The term CARM can apply to any positioner with 2 degrees of freedom of rotation of the X-Ray beam about the Imaging Subject. 2. The term COLUMN can apply to any positioner with 1 degree of freedom of rotation of the X-Ray beam about the Imaging Subject.
(0018,1510) Positioner Primary Angle
3
Position of the X-Ray beam about the patient from the RAO to LAO direction where movement from RAO to vertical is positive, if Positioner Type (0018,1508) is CARM. See C.8.7.5 XA Positioner Module for further explanation if Positioner Type (0018,1508) is CARM. See C.8.11.7 Mammography Image Module for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC.
(0018,1511) Positioner Secondary Angle
3
Position of the X-Ray beam about the patient from the CAU to CRA direction where movement from CAU to vertical is positive, if Positioner Type (0018,1508) is CARM. See C.8.7.5 XA Positioner Module for further explanation if Positioner Type (0018,1508) is CARM. See C.8.11.7 Mammography Image Module for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC.
(0018,1530) Detector Primary Angle
3
Angle of the X-Ray beam in the row direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward higher numbered columns. Negative values indicate that the X-Ray beam is tilted toward lower numbered columns. See C.8.7.5 XA Positioner Module for further explanation.
(0018,1531) Detector Secondary Angle
3
Angle of the X-Ray beam in the column direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward lower numbered rows. Negative values indicate that the X-Ray beam is tilted toward higher numbered rows. See C.8.7.5 XA Positioner Module for further explanation.
(0018,1450) Column Angulation
3
Angle of the X-Ray beam in degree relative to an orthogonal axis to the detector plane. Positive values indicate that the tilt is toward the head of the table. Note: The detector plane is assumed to be parallel to the table plane. Only meaningful if Positioner Type (0018,1508) is COLUMN.
(0018,113A) Table Type
3
Defined Terms: FIXED TILTING NONE
(0018,1138) Table Angle
3
Angle of table plane in degrees relative to horizontal plane [Gravity plane]. Positive values indicate that the head of the table is upward. Only meaningful if Table Type (0018,113A) is TILTING.
(0018,11A0) Body Part Thickness
3
The average thickness in mm of the body part examined when compressed, if compression has been applied during exposure.
(0018,11A2) Compression Force
3
The compression force applied to the body part during exposure, measured in Newtons.