DICOM Tables


PET MULTI-GATED ACQUISITION MODULE ATTRIBUTES

Tag
Tag Name
Requirement
Description
(0018,1080) Beat Rejection Flag
2
Heart beat duration sorting has been applied. Enumerated values: Y = yes N = no
(0018,1061) Trigger Source or Type
3
Text indicating trigger source. Defined terms: EKG
(0018,1085) PVC Rejection
3
Description of the type of PVC rejection criteria used.
(0018,1086) Skip Beats
3
Number of beats skipped after a detected arrhythmia.
(0018,1088) Heart Rate
3
Average number of heart beats per minute for the collection period for this image. This shall include all accepted beats as well as rejected beats.
(0018,1064) Framing Type
3
Description of type of framing performed. Defined Terms: FORW = forward BACK = backward PCNT = forward/backward by percentage