(0018,0060) |
KVP |
2
|
Peak kilo voltage output of the X-Ray generator used. |
(0018,1155) |
Radiation Setting |
1
|
Identify the general level of X-Ray dose exposure. Enumerated values are: SC = low dose exposure generally corresponding to fluoroscopic settings (e.g. preparation for diagnostic quality image acquisition); GR = high dose for diagnostic quality image acquisition (also called digital spot or cine); |
(0018,1151) |
X-Ray Tube Current |
2C
|
X-Ray Tube Current in mA. Required if Exposure (0018,1152) is not present. |
(0018,1150) |
Exposure Time |
2C
|
Duration of X-Ray exposure in msec. See 8.7.2.1.1. Required if Exposure (0018,1152) is not present. |
(0018,1152) |
Exposure |
2C
|
The exposure expressed in mAs, for example calculated from Exposure Time and X-ray Tube Current. Required if either Exposure Time (0018,1150) or X-Ray Tube Current (0018,1151) are not present. |
(0018,1153) |
Exposure in µAs |
3
|
The exposure expressed in µAs, for example calculated from Exposure Time and X-ray Tube Current. |
(0018,1166) |
Grid |
3
|
Identify the grid. Only a single value shall be present. Defined Terms are: IN = A Grid is positioned; NONE = No Grid is used. |
(0018,1154) |
Average Pulse Width |
3
|
Average width of X-Ray pulse in msec. |
(0018,115A) |
Radiation Mode |
3
|
Specifies X-Ray radiation mode. Defined Terms: CONTINUOUS PULSED |
(0018,1161) |
Type of Filters |
3
|
Type of filter(s) inserted into the X-Ray beam (e.g. wedges). |
(0018,1162) |
Intensifier Size |
3
|
Diameter of X-Ray intensifier in mm |
(0018,1147) |
Field of View Shape |
3
|
Shape of the Image Intensifier Field of View. See C.8.7.2.1.2. Defined Terms are: ROUND RECTANGLE |
(0018,1149) |
Field of View Dimension(s) |
3
|
Dimensions of the Image Intensifier Field of View in mm. If Rectangle, row dimension followed by column; if Round, diameter. |
(0018,1164) |
Imager Pixel Spacing |
3
|
Physical distance measured at the front plane of the Image Receptor housing between the center of each pixel specified by a numeric pair - row spacing value(delimiter) column spacing value in mm. |
(0018,1190) |
Focal Spot |
3
|
Nominal focal spot size in mm used to acquire this image. |
(0018,115E) |
Image Area Dose Product |
3
|
X-Ray dose, measured in dGy*cm*cm, to which the patient was exposed for the acquisition of this image plus any non-digitally recorded fluoroscopy which may have been performed to prepare for the acquisition of this image. Note: The sum of the Image Area Dose Product of all images of a Series or a Study may not result in the actual area dose product to which the patient was exposed. |