(0028,0004) |
Photometric Interpretation |
1
|
Specifies the intended interpretation of the pixel data. Shall have one of the following Enumerated Values: MONOCHROME1 MONOCHROME2 |
(0018,0060) |
KVP |
3
|
Peak kilo voltage output of the x-ray generator used |
(0018,1004) |
Plate ID |
3
|
The ID or serial number of the sensing plate upon which the image was acquired |
(0018,1110) |
Distance Source to Detector |
3
|
Distance in mm from source to detector center |
(0018,1111) |
Distance Source to Patient |
3
|
Distance in mm from source to isocenter (center of field of view) |
(0018,1150) |
Exposure Time |
3
|
Time of x-ray exposure in msec |
(0018,1151) |
X-ray Tube Current |
3
|
X-ray Tube Current in mA. |
(0018,1152) |
Exposure |
3
|
The exposure expressed in mAs, for example calculated from Exposure Time and X-ray Tube Current. |
(0018,1153) |
Exposure in µAs |
3
|
The exposure expressed in µAs, for example calculated from Exposure Time and X-ray Tube Current. |
(0018,1164) |
Imager Pixel Spacing |
3
|
Physical distance measured at the front plane of the Image Receptor housing between the center of each pixel. Specified by a numeric pair - row spacing value (delimiter) column spacing value - in mm. In the case of CR, the front plane is defined to be the external surface of the CR plate closest to the patient and radiation source. |
(0018,1170) |
Generator Power |
3
|
Power in kW to the x-ray generator. |
(0018,1400) |
Acquisition Device Processing Description |
3
|
Describes device-specific processing associated with the image (e.g. Organ Description) |
(0018,1401) |
Acquisition Device Processing Code |
3
|
Code representing the device-specific processing associated with the image (e.g. CR Organ Filtering code) |
(0018,1402) |
Cassette Orientation |
3
|
Orientation of cassette, used to properly position the image for display. Enumerated Values: LANDSCAPE PORTRAIT |
(0018,1403) |
Cassette Size |
3
|
Size of cassette. Defined Terms: 18CMX24CM 8INX10IN 24CMX30CM 10INX12IN 30CMX35CM 30CMX40CM 11INX14IN 35CMX35CM 14INX14IN 35CMX43CM 14INX17IN |
(0018,1404) |
Exposures on Plate |
3
|
Total number of x-ray exposures that have been made on the plate identified in Plate ID (0018,1004) |
(0018,1405) |
Relative X-ray Exposure |
3
|
Relative x-ray exposure on the plate. Meaning of values is implementation specific. May be used to adjust the dynamic range of the plate digitizer (scanner). |
(0018,6000) |
Sensitivity |
3
|
Read out sensitivity. |