DICOM Tables


PIXEL MEASURES MACRO ATTRIBUTES

Tag
Tag Name
Requirement
Description
(0028,9110) Pixel Measures Sequence
1
Identifies the physical characteristics of the pixels of this frame. Only a single Item shall be permitted in this sequence.
(0028,0030) >Pixel Spacing
1C
Physical distance in the patient between the center of each pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. Required if Volumetric Properties (0008,9206) is other than DISTORTED or SAMPLED. May be present otherwise.
(0018,0050) >Slice Thickness
1C
Nominal reconstructed slice thickness, in mm. See C.7.6.2.1.1 and C.7.6.16.2.3.1 for further explanation. Required if Volumetric Properties (0008,9206) is VOLUME or SAMPLED. May be present otherwise.