DICOM Tables


MODALITY LUT MODULE ATTRIBUTES

Tag
Tag Name
Requirement
Description
(0028,3000) Modality LUT Sequence
1C
Defines a sequence of Modality LUTs. Only one Item may be present. Shall not be present if Rescale Intercept (0028,1052) is present.
(0028,3002) >LUT Descriptor
1C
Specifies the format of the LUT Data in this Sequence. See C.11.1.1 for further explanation. Required if the Modality LUT Sequence (0028,3000) is sent.
(0028,3003) >LUT Explanation
3
Free form text explanation of the meaning of the LUT.
(0028,3004) >Modality LUT Type
1C
Specifies the output values of this Modality LUT. See C.11.1.1.2 for further explanation. Required if the Modality LUT Sequence (0028,3000) is sent.
(0028,3006) >LUT Data
1C
LUT Data in this Sequence. Required if the Modality LUT Sequence (0028,3000) is sent.
(0028,1052) Rescale Intercept
1C
The value b in relationship between stored values (SV) and the output units specified in Rescale Type (0028,1054). Output units = m*SV + b. Required if Modality LUT Sequence (0028,3000) is not present. Shall not be present otherwise.
(0028,1053) Rescale Slope
1C
m in the equation specified by Rescale Intercept (0028,1052). Required if Rescale Intercept is present.
(0028,1054) Rescale Type
1C