XA POSITIONER MODULE ATTRIBUTES (C.8-30)


Tag Attribute Type Description
(0018,1111)Distance Source to Patient3 Distance in mm from source to isocenter (center of field of view).
(0018,1110)Distance Source to Detector3 Distance in mm from source to detector center.
(0018,1114)Estimated Radiographic Magnification Factor3 Ratio of Source Image Distance (SID) over Source Object Distance (SOD).
(0018,1500)Positioner Motion2C Used to describe the activity of the imaging devices. Defined terms: DYNAMIC, STATIC. Required if Multi-frame data. See C.8.7.5.1.1.
(0018,1510)Positioner Primary Angle2 Position of the X-Ray Image Intensifier about the patient from the RAO to LAO direction where movement from RAO to vertical is positive. See C.8.7.5.1.2.
(0018,1511)Positioner Secondary Angle2 Position of the X-Ray Image Intensifier about the patient from the CAU to CRA direction where movement from CAU to vertical is positive. See C.8.7.5.1.2
(0018,1520)Positioner Primary Angle Increment2C Incremental change in primary positioner angle for each frame. See C.8.7.5.1.3. Required if Positioner Motion is DYNAMIC.
(0018,1521)Positioner Secondary Angle Increment2C Incremental change in secondary positioner angle for each frame. See C.8.7.5.1.3. Required if Positioner Motion is DYNAMIC.
(0018,1530)Detector Primary Angle3 Angle of the X-Ray beam in the row direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted towards higher numbered columns. Negative values indicate that the X-Ray beam is tilted towards lower numbered columns. See C.8.7.5.1.4.
(0018,1531)Detector Secondary Angle3 Angle of the X-Ray beam in the column direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted towards lower numbered rows. Negative values indicate that the X-Ray beam is tilted towards higher numbered rows. See C.8.7.5.1.4.




[ About this site ]