CR IMAGE MODULE ATTRIBUTES (C.8-2)


Tag Attribute Type Description
(0028,0004)Photometric Interpretation1 Specifies the intended interpretation of the pixel data. Shall have one of the following Enumerated Values: MONOCHROME1 MONOCHROME2
(0018,0060)KVP3 Peak kilo voltage output of the x-ray generator used
(0018,1004)Plate ID3 The ID or serial number of the sensing plate upon which the image was acquired
(0018,1110)Distance Source to Detector3 Distance in mm from source to detector center
(0018,1111)Distance Source to Patient3 Distance in mm from source to isocenter (center of field of view)
(0018,1150)Exposure Time3 Time of x-ray exposure in msec
(0018,1151)X-ray Tube Current3 X-ray Tube Current in mA.
(0018,1152)Exposure3 The exposure expressed in mAs, for example calculated from Exposure Time and X-ray Tube Current.
(0018,1153)Exposure in µAs3 The exposure expressed in µAs, for example calculated from Exposure Time and X-ray Tube Current.
(0018,1164)Imager Pixel Spacing3 Physical distance measured at the front plane of the Image Receptor housing between the center of each pixel. Specified by a numeric pair - row spacing value (delimiter) column spacing value - in mm. In the case of CR, the front plane is defined to be the external surface of the CR plate closest to the patient and radiation source.
(0018,1170)Generator Power3 Power in kW to the x-ray generator.
(0018,1400)Acquisition Device Processing Description3 Describes device-specific processing associated with the image (e.g. Organ Description)
(0018,1401)Acquisition Device Processing Code3 Code representing the device-specific processing associated with the image (e.g. CR Organ Filtering code)
(0018,1402)Cassette Orientation3 Orientation of cassette, used to properly position the image for display. Enumerated Values: LANDSCAPE PORTRAIT
(0018,1403)Cassette Size3 Size of cassette. Defined Terms: 18CMX24CM 8INX10IN 24CMX30CM 10INX12IN 30CMX35CM 30CMX40CM 11INX14IN 35CMX35CM 14INX14IN 35CMX43CM 14INX17IN
(0018,1404)Exposures on Plate3 Total number of x-ray exposures that have been made on the plate identified in Plate ID (0018,1004)
(0018,1405)Relative X-ray Exposure3 Relative x-ray exposure on the plate. Meaning of values is implementation specific. May be used to adjust the dynamic range of the plate digitizer (scanner).
(0018,6000)Sensitivity3 Read out sensitivity.




[ About this site ]
Copyright 2003 - 2016 DICOMLookup.com