Tag |
Attribute |
Type |
Description |
(0028,0004) | Photometric Interpretation | 1 | Specifies the intended interpretation of the pixel data. Shall have one of the following Enumerated Values: MONOCHROME1 MONOCHROME2 |
(0018,0060) | KVP | 3 | Peak kilo voltage output of the x-ray generator used |
(0018,1004) | Plate ID | 3 | The ID or serial number of the sensing plate upon which the image was acquired |
(0018,1110) | Distance Source to Detector | 3 | Distance in mm from source to detector center |
(0018,1111) | Distance Source to Patient | 3 | Distance in mm from source to isocenter (center of field of view) |
(0018,1150) | Exposure Time | 3 | Time of x-ray exposure in msec |
(0018,1151) | X-ray Tube Current | 3 | X-ray Tube Current in mA. |
(0018,1152) | Exposure | 3 | The exposure expressed in mAs, for example calculated from Exposure Time and X-ray Tube Current. |
(0018,1153) | Exposure in µAs | 3 | The exposure expressed in µAs, for example calculated from Exposure Time and X-ray Tube Current. |
(0018,1164) | Imager Pixel Spacing | 3 | Physical distance measured at the front plane of the Image Receptor housing between the center of each pixel. Specified by a numeric pair - row spacing value (delimiter) column spacing value - in mm. In the case of CR, the front plane is defined to be the external surface of the CR plate closest to the patient and radiation source. |
(0018,1170) | Generator Power | 3 | Power in kW to the x-ray generator. |
(0018,1400) | Acquisition Device Processing Description | 3 | Describes device-specific processing associated with the image (e.g. Organ Description) |
(0018,1401) | Acquisition Device Processing Code | 3 | Code representing the device-specific processing associated with the image (e.g. CR Organ Filtering code) |
(0018,1402) | Cassette Orientation | 3 | Orientation of cassette, used to properly position the image for display. Enumerated Values: LANDSCAPE PORTRAIT |
(0018,1403) | Cassette Size | 3 | Size of cassette. Defined Terms: 18CMX24CM 8INX10IN 24CMX30CM 10INX12IN 30CMX35CM 30CMX40CM 11INX14IN 35CMX35CM 14INX14IN 35CMX43CM 14INX17IN |
(0018,1404) | Exposures on Plate | 3 | Total number of x-ray exposures that have been made on the plate identified in Plate ID (0018,1004) |
(0018,1405) | Relative X-ray Exposure | 3 | Relative x-ray exposure on the plate. Meaning of values is implementation specific. May be used to adjust the dynamic range of the plate digitizer (scanner). |
(0018,6000) | Sensitivity | 3 | Read out sensitivity. |