US IMAGE MODULE ATTRIBUTES (C.8-18)


Tag Attribute Type Description
(0028,0002)Samples Per Pixel1 Number of samples (planes) in this image. See C.8.5.6.1.12 for specialization
(0028,0004)Photometric Interpretation1 Specifies the intended interpretation of the pixel data. See C.8.5.6.1.2 for specialization.
(0028,0100)Bits Allocated1 Number of bits allocated for each pixel sample. See C.8.5.6.1.13 for specialization.
(0028,0101)Bits Stored1 Number of bits stored for each pixel sample. See C.8.5.6.1.14 for specialization.
(0028,0102)High Bit1 Most significant bit for pixel sample data. See C.8.5.6.1.15 for specialization.
(0028,0006)Planar Configuration1C Indicates whether the pixel data are sent color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a value greater than 1. See C.8.5.6.1.16 for specialization.
(0028,0103)Pixel Representation1 Data representation of pixel samples. See C.8.5.6.1.3 for specialization.
(0028,0009)Frame Increment Pointer1C Contains the Data Element Tag of the attribute which is used as the frame increment in Multi-frame pixel data (see C.7.6.6). Required if Number of Frames is sent. See C.8.5.6.1.4 for specialization.
(0008,0008)Image Type2 Image identification characteristics. See C.8.5.6.1.1 for specialization.
(0028,2110)Lossy Image Compression1C Specifies whether an Image has undergone lossy compression. Enumerated Values: 00 = Image has NOT been subjected to lossy compression. 01 = Image has been subjected to lossy compression. See C.7.6.1.1.5 Required if Lossy Compression has been performed on the Image.
(0008,2124)Number of Stages2C Number of Stages in this protocol. Required if image was acquired in a Staged protocol.
(0008,212A)Number of Views in Stage2C Number of views in this Stage. Required if image was acquired in a Staged protocol.
(0028,0014)Ultrasound Color Data Present3 This element indicates if any ultrasound color data is present in an image. Enumerated Values: 00 = Ultrasound color data not present in image 01 = Ultrasound color data is present in image. See C.8.5.6.10
(0008,1130)Referenced Overlay Sequence3 A Sequence which provides reference to a set of SOP Class/Instance pairs which are related independent Overlays. Uniquely identifies Overlays significantly related to this Image. Zero or more Items may be included in this Sequence.
(0008,1150)>Referenced SOP Class UID1C Uniquely identifies the referenced SOP Class. Required if Referenced Overlay Sequence (0008,1130) is sent.
(0008,1155)>Referenced SOP Instance UID1C Uniquely identifies the referenced SOP Instance. Required if Reference Overlay Sequence (0008,1130) is sent. Referenced Curve Sequence (0008,1145) 3 A sequence which provides reference to a set of SOP Class/Instance pairs which are related independent Curves. Uniquely identifies Curves significantly related to this Image.
(0008,1150)>Referenced SOP Class UID1C Uniquely identifies the referenced SOP Class. Required if Referenced Curve Sequence (0008,1145) is sent.
(0008,1155)>Referenced SOP Instance UID1C Uniquely identifies the referenced SOP Instance. Required if Reference Curve Sequence (0008,1145) is sent.
(0008,2120)Stage Name3 A Stage is a particular time slice of a protocol in which a set of images are collected. The names can be free form text. Recommended text for Stress Echo stage names are: PRE-EXERCISE, POST-EXERCISE, PEAK-EXERCISE, RECOVERY, BASELINE, LOW DOSE, PEAK DOSE
(0040,000A)Stage Code Sequence3 Sequence of items describing the performed Ultrasound Protocol Stage(s).
(0008,2122)Stage Number3 A number that identifies the Stage. Stage Number starts at one.
(0008,2127)View Name3 A View is a particular combination of the position and orientation when a set of images are acquired. Images are acquired at the same View in different Stages for the purpose of comparison.
(0008,2128)View Number3 A number that identifies the View. View Number starts at one.
(0008,2129)Number of Event Timers3 The number of event timers used at the time of acquisition of a Multi-frame image.
(0008,2130)Event Elapsed Time(s)3 An array of values associated with each event timer. Units in milliseconds.
(0008,2132)Event Timer Name(s)3 Name that identifies the event timer.
(0008,2218)Anatomic Region Sequence3 Sequence of one Item that identifies the anatomic region of interest in this image (i.e. external anatomy, surface anatomy, or general region of the body). See Section C.8.5.6.1.17.
(0008,2220)>Anatomic Region Modifier Sequence3 Sequence of one or more Items that modifies the anatomic region of interest in this image (i.e. prone, supine, decubitus right). See Section C.8.5.6.1.17.
(0008,2228)Primary Anatomic Structure Sequence3 Sequence of one or more Items that identifies the primary anatomic structure of interest in this image. See Section C.8.5.6.1.18
(0008,2230)>Primary Anatomic Structure Modifier Sequence3 Sequence of one or more Items that modifies the primary anatomic structure of interest in this image. See Section C.8.5.6.1.18
(0008,2240)Transducer Position Sequence3 Sequence of one or more Items that identifies the transducer position used in this image. See Section C.8.5.6.1.19.
(0008,2242)> Transducer Position Modifier Sequence3 Sequence of one or more Items that modifies the primary transducer position of interest in this image. See Section C.8.5.6.1.19.
(0008,2244)Transducer Orientation Sequence3 Sequence of one or more Items that identifies the Transducer Orientation used in this image. See Section C.8.5.6.1.20.
(0008,2246)> Transducer Orientation Modifier Sequence3 Sequence of one or more Items that modifies the primary Transducer Orientation of interest in this image. See Section C.8.5.6.1.20
(0008,002A)Acquisition Datetime1C The date and time that the acquisition of data that resulted in this image started. Required if Modality (0008,0060) = IVUS May be present otherwise. Note: The synchronization of this time with an external clock is specified in the Synchronization Module in Acquisition Time Synchronized (0018,1800).
(0018,1060)Trigger Time3 Time interval measured in msec from the start of the R-wave to the beginning of data taking
(0018,1062)Nominal Interval3 Average R-R interval used for these data, in msec
(0018,1080)Beat Rejection Flag3 Beat length sorting has been applied. Enumerated Values: Y = yes N = no
(0018,1081)Low R-R Value3 R-R interval low limit for beat rejection, in msec
(0018,1082)High R-R Value3 R-R interval high limit for beat rejection, in msec
(0018,1088)Heart Rate3 Beats per minute.
(0018,3100)IVUS Acquisition1C Defined Terms: MOTOR_PULLBACK MANUAL_PULLBACK SELECTIVE GATED_PULLBACK See C.8.5.6.1.21 Required if Modality (0008,0060) = IVUS
(0018,3101)IVUS Pullback Rate1C Required if IVUS Acquisition (0018,3100) value is MOTOR_PULLBACK. Specified in units of mm/sec. See C.8.5.6.1.22
(0018,3102)IVUS Gated Rate1C Required if IVUS Acquisition (0018,3100) value is GATED_PULLBACK. Specified in units of mm/beat. See C.8.5.6.1.23
(0018,3103)IVUS Pullback Start Frame Number1C Required if IVUS Acquisition (0018,3100) value is MOTOR_PULLBACK or GATED_PULLBACK. See C.8.5.6.1.24
(0018,3104)IVUS Pullback Stop Frame Number1C Required if IVUS Acquisition (0018,3100) value is MOTOR_PULLBACK or GATED_PULLBACK. See C.8.5.6.1.25
(0018,3105)Lesion Number3 Identifier(s) of the lesion(s) of interest imaged within the current SOP Instance. Each lesion shall have a unique numeric integer identifier within the study. See C.8.5.6.1.26.
(0018,5000)Output Power3 Manufacturer defined character string description of ultrasound output level(s) used in generating a given image. Data may be expressed in dB, %, W/cm 2 , etc.
(0018,5010)Transducer Data3 Manufacturer defined code or description of ultrasound transducer used.
(0018,6031)Transducer Type3 Defined Terms: SECTOR_PHASED SECTOR_MECH SECTOR_ANNULAR LINEAR CURVED LINEAR SINGLE CRYSTAL SPLIT XTAL CWD IV_PHASED IV_ROT XTAL IV_ROT MIRROR ENDOCAV_PA ENDOCAV_MECH ENDOCAV_CLA ENDOCAV_AA ENDOCAV_LINEAR VECTOR_PHASED
(0018,5012)Focus Depth3 The depth, from the transducer face, of the manufacturer defined beam focus used for the image, in cm.
(0018,5020)Preprocessing Function3 Manufacturer defined description of processing of echo information. Data may include code or description of gain (initial, overall, TGC, dynamic range, etc.), preprocessing, postprocessing, Doppler processing parameters, e.g. cutoff filters, etc., as used in generating a given image.
(0018,5022)Mechanical Index3 See C.8.5.6.1.8 for Description.
(0018,5024)Bone Thermal Index3 See C.8.5.6.1.8 for Description.
(0018,5026)Cranial Thermal Index3 See C.8.5.6.1.8 for Description.
(0018,5027)Soft Tissue Thermal Index3 See C.8.5.6.1.8 for Description.
(0018,5028)Soft Tissue-focus Thermal Index3 See C.8.5.6.1.8 for Description.
(0018,5029)Soft Tissue-surface Thermal Index3 See C.8.5.6.1.8 for Description.
(0018,5050)Depth of Scan Field3 The depth, in mm, from the transducer face to the deepest point included in the displayed image– the field of view.
(0018,5210)Image Transformation Matrix3 Transformation Matrix, as described in C.8.5.6.1.9. Used for spatially related images where the relationship includes rotations.
(0018,5212)Image Translation Vector3 Translation Vector as described in C.8.5.6.1.9. Used for spatially related images where the relationship includes translations.
(60xx,0045)Overlay Subtype3 Defined term which identifies the intended purpose of the ROI Overlay Type. See C.8.5.6.1.11 for specialization.