DICOM Tables


Waveform Annotation Module Attributes

Tag
Tag Name
Requirement
Description
(0040,B020) Waveform Annotation Sequence
1
Sequence of Annotation Items; one or more items shall be present
(0070,0006) > Unformatted Text Value
1C
Text Observation Value (annotation). Mutually exclusive with Concept Name Code Sequence (0040,A043)
(0040,A043) > Concept Name Code Sequence
1C
Code representing the fully specified name of the NUMERIC measurement or CODED concept. This sequence shall contain exactly one item. Mutually exclusive with Text Value (0070,0006).
(0040,A195) >> Modifier Code Sequence
1C
A sequence of items modifying or specializing the Concept Name. Any number of items may be present. Required if Concept Name Code Sequence (0040,A043) is sent and the value does not fully describe the semantics of the measurement or concept.
(0040,A168) > Concept Code Sequence
3
A sequence that conveys the categorical coded nominal value.
(0040,A195) >> Modifier Code Sequence
1C
A sequence of items modifying or specializing the Concept. Any number of items may be present. Required if Concept Code Sequence (0040,A168) is sent and the value does not fully describe the semantics of the concept value.
(0040,A30A) > Numeric Value
3
Numeric measurement value or values.
(0040,08EA) > Measurement Units Code Sequence
3
Units of measurement. Coded entry sequence with one item only.
(0040,A0B0) > Referenced Waveform Channels
1
List of channels in waveform to which annotation applies. See C.10.10.1.1
(0040,A130) > Temporal Range Type
1C
See C.10.10.1.2 for Enumerated Values. Required if Annotation does not apply to entire Referenced Waveform Channels; shall not be present if Annotation applies to entire temporal extent of referenced channels.
(0040,A132) > Referenced Sample Positions
1C
List of samples within a multiplex group specifying temporal points for annotation. Position of first sample is 1. Required if Temporal Range Type (0040,A130) is present, and if Referenced Time Offsets (0040,A138) and Referenced Datetime (0040,A13A) are not present. See C.10.10.1.3
(0040,A138) > Referenced Time Offsets
1C
Specifies temporal points for annotation by number of seconds after start of data. Required if Temporal Range Type (0040,A130) is present, and if Referenced Sample Positions (0040,A132) and Referenced Datetime (0040,A13A) are not present.
(0040,A13A) > Referenced Datetime
1C
Specifies temporal points for annotation by absolute time. Required if Temporal Range Type (0040,A130) is present, and if Referenced Sample Positions (0040,A132) and Referenced Time Offsets (0040,A138) are not present.
(0040,A180) > Annotation Group Number
3